Specifications:
NIBP:
Patient: Adult and Pediatric
Measurement method: Oscillometric
Range:Adult:30mmHg~255mmHg Pediatric: 30mmHg~160mmHg
Measurement unite: mmHg/kPa selectable
Measurement mode: Manual, Auto, STAT
Measurement Type: SYS, DIA, MAP
Measurement interval in AUTO mode:1,2,3,5,10,15,20,30,45,60Min
Measuring period in STAT mode: 5 Min
Resolution:1mmHg
Over-pressure protection:Adult:300mmHg Pediatric:220mmHg
SPO2:
Patient: Adult, child
Range: 0 -100%
Resolution: 1 %
Accuracy: ±2 %(90~99%), ±3 %(70~89%),unspecified (<70%)
PR:
Range:30bpm~250bpm
Resolution:1bpm
Accuracy: ±2bpm
Temperature:
Channel: 1
Input: Body surface thermal-sensitive resistor temperature sensor
Measurement Unit: °C/°F selectable
Range: 0~50℃
Resolution: 0.1℃
Accuracy: ±0.1℃
Environment:
Temperature
Operation:0°C~45°C
Transportation and storage:-20°C~60°C
Humidity
Operation: 15%~95 %( non-condensing)
Transportation and storage: 10%~95% (non-condensing)
Altitude
Operation:86kPa ~106kPa
Transportation and storage:50kPa~106kPa
Power Requirements:
AC power adaptor
Input: AC 100~240V, 50/60Hz, 0.7A
Output: DC 8.4V,1A
Internal battery
Type: 1500mAh/7.2V Li-ion rechargeable battery
Operating Time: 5 hours of stand-by time with a full battery.
Recharge time: 8 hours