Features and Specifications

Specifications 

Fluoroscopic Capacity  
 Max rated capacity           : Tube Current 4mA, Tube Voltage 110kV  
 Automatic Fluoroscopy     : Tube Voltage:40kV110kV adjust 
                                              automatically 
                                   Tube Current  : 0.3mA4mA set manually  
 Manual Fluoroscopy          : Continuous Tube Voltage:40kV110kV 
                        Continuous Tube Current : 0.3mA4mA  
Pulse Fluoroscopy             : Continuous Tube Voltage 40kV110kV  
                                              Continuous Tube Current 4.1mA8mA   
Photography Capacity  
 Max rated capacity             : 3.5 KW  
 Tube Voltage and Cur- 
   rent Combination               : 40kV49kV   1 125 mAs 
                                                50kV59kV   1 110 mAs 
                                                60kV69kV   1 90  mAs 
                                                70kV79kV   1 80  mAs 
                                                80kV89kV   1 71  mAs 
                                                90kV99kV   1 63  mAs 
                                                100kV110kV 1 40  mAs  
 Plateholder Size                 :  200mm×250mm(8″×10″)or  
                                               250mm×300mm10″×12″  
X-ray Tube  
   X-ray Tube Special for High Frequency

 Fixed anode   Dual-focus : large focus: 1.5, small focus: 0.6 

Inverter Frequency:  40KHz 

Thermal capacity: 35KJ (47KHU)  
VideoSystem  
   Image Intensifier       : Image Intensifier made by TOSHIBA (9″、 
                                       6″4.5″)  
   CCD Vidicon              : Ultra low-light CCD camera imported from Japan                              
   Monitor                       : Horizontal 1000 lines and vertical 800 lines, 
                                         Bandwidth : 12.5MHz,     Image/sec : 75
                                

CCU(central control)         : Recursive Filter: K=8, 8 images storage,  
                                               image upright, image overturnpositive &  
                                               negative imageLIH(last image freeze, and 
                                               OSDmonitor display  
Structure : 
 Directive wheel                  : ±90°revolutioncan freely change the moving  
                                              direction of the unit.
 Ascending & Descend- 
 ing Range of Pillar             : ≥400mm  
 C-arm                                : Forward and Backward Movement: 200mm 
                                             Revolution around Horizontal Axis:  ±180° 
                                              Revolution around Vertical Axis     :  ±15° 
                                              Slip on orbit: 120°(+90° -30°)  

Features 

1With a compact appearance, and easy to operate; 
2Clinical Video system with high performance8 images storage 
       volume, and two 14high-resolution monitors; 
3Automatically track fluoroscopy to make the image brightness and  
      clearness optimum; 
4Automatically store the last image of fluoroscopy to advantage 
      diagnosis; 
5With a high-quality knockdown X-ray generator to reduce radiation 

Clause/ Specification

Section A 

Main Digital Radiography System

1.0

X-ray Generator 

Each set of X-ray generator unit for general Fluoroscopy and Radiography with at least the following features:

1.1

3-phase ,high frequency converter / inverter, multi-pulse and microprocessor control 

1.2

Nominal power output 3.5 KW

1.3

Tube Voltage :40KV~110KV, Tube current output of 20~63mA ,1.0 mAs~125mAs

1.4

Exposure time range :from 2 ms or shorter to 2 seconds or longer

1.5

Suitable to operation of dual-focus X-ray tube : Large Focus :1.5mm,Small Focus :0.6 mm

1.6

Automatic line voltage compensation

1.7

Automatic exposure control with manual override 

1.8

Digital display of operator pre-set exposure data

1.9

Digital display of actual exposure data after X-ray exposure

1.10

Self-diagnostic functions with digital display of error codes

1.11

Tube-overload indication and protection on the control panel

1.12

User programmed radiographic exposure techniques with the following features:

1.13

Unique double foot brake controller design makes it convenient to control the instrument inside and outside the operation room. This design can also protect the health care professionals by reducing the opportunity to come into contact with the radiation. 

1.14

Each exposure programme includes at least 2 additional selections for appropriate patient size compensation ( normal figure, fat figure,thin figure, adult, child , normal position lateral position )

1.15

The exposure programmes are user programmable

1.16

Selection of exposure programmes are by touch-controls 

1.17

Alphanumeric display of selected anatomical exposure programme

1.18

Integrated radiation exposure switch on generator control console

1.19

All necessary protective circuits are incorporated .

2.0

X-ray Tube

Each set of rotation anode X-ray tube to be mounted to the Tube Carries System with at least the following features: 

2.1

Dual focus with focal spot sizes of 0.6 mm and  focus 1.5mm 

2.2

Tube voltage 110 kV

2.3

Anode heat storage capacity is 35 KJ 

2.4

Safety circuitry to prevent over heating with indication on control console of X-ray generator

3.1

Collimator

Each set of light beam collimator to is mounted onto monoblock with below features :Manually adjustable diaphragms with indicator/readings of beam coverage at selected Source-Image-Distance(SID) 

3.2

Automatic collimation according to examination selected, can be overridden manually 

3.3

Optical centering indication for positioning and side beam for central alignment of X-ray beam with detector assembly

3.4

Receptacle channels(s)/track(s) for insertion of external filters. A full set of dedicated filters, with filter holder for attachment to the collimator, has to be provided for exposure compensation for various anatomical regions such as skull , shoulder, spine and chest etc

3.5

built-in electronic or mechanical measuring device for measurement of SID is incorporated 

3.6

Provision of dose reduction filters , which are operator selectable or automatic for filtering of low energy radiation

3.7

Integrated display/multi-functional control panel with at least the following features:

3.8

Selection of target detector :

Table 

Wall stand 

Table top for free exposure

4.1

All movements of the tube assembly can be manually operated and are provided with electro-magnetic brakes.

4.2

Automatic height adjustment of X-ray tube to maintain constant focus to tabletop distance when table height changes.

4.3

Automatic centering of X-ray tube to remain centre to the vertical detector when the height of detector assembly changes 

4.4

Remote control of automatic detector tracking 

4.5

It can be controlled at the acquisition workstation or at other convenient facility such as hand control .

4.6

Safety mechanism to prevent the X-ray tube or the column assembly from failing in case of lock, springs or cables failure.

4.7

C arm Revolution around Vertical axis: ±15°, Horizontal axis: ±180°,

5.0

PLX150 Mobile Surgical Table for C arm with the following features:

5.1

The surface of table adopt pneumatic support, hands lifting, reliable operation and comfortable. It is not only absorb small dose of X-ray but also the uniform of X ray .

5.2

The both ends of the table surface can be lifted respectively to tilt the table surface to meet the needs of special examination.

5.3

The table surface can be separated from the table body,  reducing the number of patient handling, thus conducive to the rehabilitation of postoperative

5.4

Table-top movement range

5.5

Longitudinal: ±250mm, Size of the table is 1960x586 mm. Patient weight carrying capacity of tabletop is 135kg  Lowest table height,:95KG

5.6

Integrated with the CCD detector 

5.7

Equipped with a removable, oscillating grid ,grid ratio 10:1 

5.8

Bucky assembly have a vertical movement range of 470mm~1468mm

6.0

The table detector shall have the following features:

6.1

Data acquisition is 14 bits per pixel

7.0

Image Acquisition console

One set of Image Acquisition Console for each unit of Digital Radiography Imaging System on offer is provided and installed with the following performance for X-ray control ,image acquisition, processing and display 

7.1

High-end PC processor with network capability . PC processor of highest processing capacity .

7.2

It can be connected to and control the operation of the generator system and be able to control the image acquisition parameters

7.3

The image Acquisition Console support DICOM 3.0, Able to accept accession number from PACS,Dicom DIR,Film printer, mass storage ,  DICOM 3.0 with conformance statements DICOM Send,  DICOM Print, DICOM Storage, 

7.4

Different formats of images can be selected for printing 

7.5

14 LCD monitor , of resolution not less than 1280 x 1024,for image display . Keyboard ,mouse and barcode reader are provided for data entry. 

Product Name:
High Frequency Mobile C arm x ray  
Item No: TDM8112