Fluoroscopic Capacity
Max rated capacity : Tube Current 4mA, Tube Voltage 110kV
Automatic Fluoroscopy : Tube Voltage:40kV~110kV adjust
automatically
Tube Current : 0.3mA~4mA set manually
Manual Fluoroscopy : Continuous Tube Voltage:40kV~110kV
Continuous Tube Current : 0.3mA~4mA
Pulse Fluoroscopy : Continuous Tube Voltage 40kV~110kV
Continuous Tube Current 4.1mA~8mA
Photography Capacity
Max rated capacity : 3.5 KW
Tube Voltage and Cur-
rent Combination : 40kV~49kV 1 ~125 mAs
50kV~59kV 1 ~110 mAs
60kV~69kV 1 ~90 mAs
70kV~79kV 1 ~80 mAs
80kV~89kV 1 ~71 mAs
90kV~99kV 1 ~63 mAs
100kV~110kV 1 ~40 mAs
Plateholder Size : 200mm×250mm(8″×10″)or
250mm×300mm(10″×12″)
X-ray Tube
X-ray Tube Special for High Frequency
Fixed anode Dual-focus : large focus: 1.5, small focus: 0.6
Inverter Frequency: 40KHz
Thermal capacity: 35KJ (47KHU)
VideoSystem
Image Intensifier : Image Intensifier made by TOSHIBA (9″、
6″、4.5″)
CCD Vidicon : Ultra low-light CCD camera imported from Japan
Monitor : Horizontal 1000 lines and vertical 800 lines,
Bandwidth : 12.5MHz, Image/sec : 75
CCU(central control) : Recursive Filter: K=8, 8 images storage,
image upright, image overturn,positive &
negative image;LIH(last image freeze, and
OSD(monitor display)
Structure :
Directive wheel : ±90°revolution,can freely change the moving
direction of the unit.
Ascending & Descend-
ing Range of Pillar : ≥400mm
C-arm : Forward and Backward Movement: 200mm
Revolution around Horizontal Axis: ±180°
Revolution around Vertical Axis : ±15°
Slip on orbit: 120°(+90°~ -30°)
|
I Usage:
Orthopaedics: osteopathy, diaplasis, nailing
Surgery: removing foreign body, cardiac catheter, implanting pace maker, interventional therapy, partial radiography, local photography, and other work.
II Features
1. New integrated design, compact appearance, convenient operation.;
2. Unique hand-held remote controller, more convenient operation
3. Graphical colorfully LCD touch screen console interface, convenient operation, fashionable design
4. High-quality combined HF-HV X-ray generator, greatly reducing the X-ray irradiation.
5. Toshiba image intensifier is employed, stable and reliable in quality , high-definition image and display system
6. With the perspective KV, MA automatic tracking feature, the image brightness and definition are automatically set at optimal function.
7. Medical high-resolution progressive output TV system, eight image storage, clearer image and convenient operation.
III Configuration
1, New (with electric auxiliary support arm) C-arm main rack one set
2, High-frequency high-voltage X-ray generator and high-frequency inverter power supply(3.5KW、40KHZ、110KV) one set
3, Toshiba 9-inch image intensifier one set
4, 1 million pixel ultra-low-light digital cameras one set
5, high- definition progressive output CCU main-control system one set
6, High-thickness grids one set
7, Electric adjustable collimator one set
8, 14-inch high- definition medical monitor one set
9, Unique hand-held remote controller one set
Ⅵ System Performance and Parameter
Category
|
Project
|
Content
|
Electrical
Performance
|
Photography Capacity
|
3.5KW
|
Fluoroscopy Capacity
|
Tube current:8mA Tube Voltage:110KV
|
Automatic Fluoroscopy
|
Tube Voltage:40kV~110kV automatical adjustment
Tube Current:0.3mA~4mA automatical adjustment
|
Manual Fluoroscopy
|
Tube Voltage:40kV~110kV Continuous
Tube Current:0.3mA~4mA Continuous
|
Pulse Fluoroscopy
|
Tube Voltage:40kV~110kV Continuous
Tube Current:4.1mA~8mA Continuous
|
Photography Tube Voltage and mAs
|
40 kV~110 kV 20-63mA 1-125mAs
|
film holder Size
|
200mm×250mm(8″×10″)or
250mm×300mm(10″×12″)
|
X-ray Tube
|
X-ray Tube Special for High Frequency
|
Fixed anode Dual-focus, small focus: 0.3, large focus:1.5 High Frequency Inverter: 40KHz thermal capacity 35KJ (47KHU) Tube thermal capacity 650KJ(867KHU)
|
Video
System
|
Image Intensifier
|
TOSHIBA (9″) Image Intensifier
|
CCD Video camera
|
Ultra low-light CCD camera
|
Monitor
|
Horizontal center resolution 1000/L, frequency for picture refresh is 75HZ
|
CCU main control
|
High Definition progressive output mode, 8 level Noise Reduction,8 image storage, Vertical Image, Horizontal image flip, positive or negative image
|
Structure
|
Directive wheel and major wheel
|
Directive wheel 360° rotation, freely change the moving direction of the unit, major wheel ±90° rotation
|
C-arm
|
Electrical Ascending & Descend ≥400mm
Monitor rotation 135°
Forward and Backward Movement: 200mm
Rotation around Horizontal Axis: ±180°
Rotation around Vertical Axis : ±15°
Focus screen distance:960mm C-arm opening: 760mm
C-arm arc depth: 640mm Slip on orbit: 120°(+90°~ -30°)
|
|