Features and Specifications

Specifications 

Power Output                       : 25KW 
Inverter Frequency                  : 40kHz 
X-Ray Tube                          :  XD56-5  17/130 
Dual-focus                    :  Small focus: 0.6     Large focus: 1.3

Thermal capacity : 212kJ (300 KHU) 
The speed of the rotary anode     : 3000rpm 
Tube Voltage                        : 40~125kV 
Tube Current                        : 200mA 
mAs                                  : 0.4~360mAs 
Time of exposure 0.04~2ms ,through adjust mAs to modify ms by 54 steps

High Voltage rectifier system, three phase full wave bridge rectifying .
Table radiography : 

Grid density :103L/INCH;  Grid ratio:8:1 

Focusing distance:120cm  Stationary type: 15"×18"(import) 

Bucky stand radiography: 

Grid density :103L/INCH;  Grid ratio:10:1 

Focusing distance:150cm Stationary type:18"×18"(import) 

Bucky 80KV Hour HVL≥2.3mmAl, 120kV Hour HVL≥3.2mmAl

Power supply : 380V±10%  50Hz, Capacity: ≥30KVA
Operation method : close-table graphic program controlled LCD touch  screen operation ;wireless remote control exposure 

(Standard Configuration: control console of close-table graphic program controlled LCD touch screen operation)

Spare parts: Protect Ground cable , 8m*1 , 12m*1 ; B. Filter :1mm*1 ,2mm*1 ; C. Remote Control *1 ; D.Accessory case*1 set; E. Hand-held switch*1 set 

Features:

1. It is installed with close-table graphical colorful LCD touch screen, wireless remote control exposure, make the operator easy to use..

2. We incorporate the latest compact high-frequency and high-voltage X-ray generator, which ensure superior image quality, while the skin dose is low , to protect the patient better.

3. Adopt the technology of kV and mAs numeral Closed loop control , Real-time control of the microprocessor , which ensure the accuracy and repeatability of the output dose.

4. Be equipped with radiography parameter of multi-part, multi-posture, multi-body type’s  adult ,child physical characteristic, the operator can modify and save the parameter by himself, which make the operation more conveniently. 

5. Possess multiple automatic protection and fault clue function, make the operation more safe and reliable.

6. The radiography table can float in any direction, electromagnetic braking, position the patient more convenience and accuracy. 

7. X-ray generator component can move round the axis of X-ray generator, crossarm and Pillar stand , be convenient with side and slanting direction projection and radiography of patient on the stretcher.

8. Be equipped with the bucky stand, which is used in the radiography of the head, chest, belly, pelvic cavity, backbone and periphery.

9. Control console of graphic program controlled LCD touch screen is optional, which can set the radiography parameter and exposure in another room.

Ⅲ Specification:

Item

content

Power Output

25KW

Inverter Frequency 

40KHz

X-ray Tube 

Rotary anode, Dual-focus, Large focus: 1.3mm, 
Small focus: 0.6, Thermal capacity:900kJ 

The speed of the rotary anode

3000rpm

Tube Current 

200mA

Tube Voltage 

40-125kV

mAs

0.4-360mAs

Table radiography

Grid density:103L/INCH;Grid ratio:10:1;Focusing distance:120cm;Stationary type:15"×18"(import)

Bucky stand radiography

Grid density:103L/INCH;Grid ratio:10:1;Focusing distance:150cm;Stationary type:18"×18"(import)

Power Supply

380V  50Hz, Capacity: ≥30KVA

Operation Method

close-table graphic program controlled LCD touch screen operation;wireless remote control exposure 

Ⅳ Basic Configuration:

1, Novel integrative planigraphy couch                                        1 uint

2, Compact high-frequency and high-voltage X-ray generator                     1 unit

3, Coloured LCD touch screen control box                                     1 unit     

4, Control system (with software)                                             1 unit

5, Bucky stand (with the South Korean Rulers filter line devices )                  1 unit

Upright Bucky Stand 

The height : 1768 mm

Movement range of up and down : 470mm~1468 mm

Grid Density :103L/INCH 

Grid Ratio : 10:1 

Focus Distance : 150cm ,

Features:  with Stationary Filter line, 18 X18 

         Can take maximum 17X17 X ray film  

Ⅴ Structural performance 

Size of the tabel 

2000×760mm

Height of the table

 ≤700mm

Transverse travel of table

±110mm

Longitudinal travel of table   

±900mm

Movement of the Grids cassette holder

≥560mm

X-ray tube assembly center rotation

±90°

Rotation of the pillar

4×90°

Axial rotation

0-35°

SID

450-1200mm

Pillar moving along the table

≥1350mm

Limited time of analog light beam opening

Approx 30s

Size of the film

5"×7"-14"×17"(Table radiography)

Size of the film

5"×7"-17"×17"(Bucky stand radiography)

Fluctuate Move Range of Chest X-ray frame

560-1625mm

Ⅵ Feature:

1. It is installed with close-table graphical colorful LCD touch screen, wireless remote control exposure, make the operator easy to use. 

2. We incorporate the latest compact high-frequency and high-voltage X-ray generator, which ensure superior image quality, while the skin dose is low , to protect the patient better. 

3. Adopt the technology of kV and mAs numeral Closed loop control , Real-time control of the microprocessor , which ensure the accuracy and repeatability of the output dose. 

4. Be equipped with radiography parameter of multi-part, multi-posture, multi-body type’s? adult ,child physical characteristic, the operator can modify and save the parameter by himself, which make the operation more conveniently. 

5. Possess multiple automatic protection and fault clue function, make the operation more safe and reliable. 

6. The radiography table can float in any direction, electromagnetic braking, position the patient more convenience and accuracy. 

7. X-ray generator component can move round the axis of X-ray generator, crossarm and Pillar stand , be convenient with side and slanting direction projection and radiography of patient on the stretcher. 

8. Be equipped with the bucky stand, which is used in the radiography of the head, chest, belly, pelvic cavity, backbone and periphery. 

9. Control console of graphic program controlled LCD touch screen is optional, which can set the radiography parameter and exposure in another room. 

Ⅶ Main Parameter 

Clause/ Specification

Section A 

Main Digital Radiography System

1.0

X-ray Generator 

Each set of X-ray generator unit for general radiography with at least the following features:

1.1

3-phase ,high frequency converter / inverter, multi-pulse and microprocessor control 

1.2

Nominal power output 15KW

1.3

Tube current output of 200mA ,0.4mAs~360mAs

1.6

Exposure time range :from 2 ms or shorter to 2 seconds or longer

1.7

Suitable to operation of dual-focus X-ray tube : Large Focus :0.6 mm,Small Focus :0.3 mm

1.8

Solid-state rectification

1.9

Automatic line voltage compensation

1.10

Automatic exposure control with manual override 

1.11

Digital display of operator pre-set exposure data

1.12

Digital display of actual exposure data after X-ray exposure

1.13 

Self-diagnostic functions with digital display of error codes

1.14 

Tube-overload indication and protection on the control panel

1.15

User programmed radiographic exposure techniques with the following features:

1.15.1

To be provided with at least 120 organ-related anatomical programmes to cover the whole body  

1.15.2

Each exposure programme include at least 2 additional selections for appropriate patient size compensation 

1.15.3

The exposure programmes is user programmable

1.15.4

Selection of exposure programmes are by touch-controls or push button controls 

1.15.5

Alphanumeric display of selected anatomical exposure programme

1.15.6

Integrated radiation exposure switch on generator control console

1.17

All necessary protective circuits are incorporated .

2.0

X-ray Tube

Each set of rotation anode X-ray tube to be mounted to the Tube Carries System with at least the following features: 

2.1

Dual focus with focal spot sizes of 0.6 mm and  focus 0.3mm 

2.2

Tube voltage 125 kV

2.3

Anode heat storage capacity is 300 KHU 

2.4

Safety circuitry to prevent over heating with indication on control console of X-ray generator

3.1

Collimator

Each set of light beam collimator to is mounted onto the Tube Unit with below features :Manually adjustable diaphragms with indicator/readings of beam coverage at selected Source-Image-Distance(SID) 

3.2

Automatic collimation according to examination selected, can be overridden manually 

3.3

Optical centering indication for positioning and side beam for central alignment of X-ray beam with detector assembly

3.4

Receptacle channels(s)/track(s) for insertion of external filters. A full set of dedicated filters, with filter holder for attachment to the collimator, has to be provided for exposure compensation for various anatomical regions such as skull , shoulder, spine and chest etc

3.5

built-in electronic or mechanical measuring device for measurement of SID is incorporated 

3.6

Provision of dose reduction filters , which are operator selectable or automatic for filtering of low energy radiation

3.7

Integrated display/multi-functional control panel over X-ray tube housing with at least the following features:

4.1

All movements of the tube assembly can be manually operated and are provided with electro-magnetic brakes.

4.2

Automatic centering of X-ray tube to remain centre to the vertical detector when the height of detector assembly changes 

4.3

Automatic longitudinal detector tracking.

4.4

Remote control of automatic detector tracking 

4.5

Auto-positioning function shall be provided

4.6

It can be controlled at the acquisition workstation or at other convenient facility such as hand control .

4.7

Safety mechanism to prevent the X-ray tube or the column assembly from failing in case of lock, springs or cables failure.

4.8

Column rotation angle :4*90°(360°), Monoblock rotation angle: +90°/-90°,

X ray tube axial rotation: 0~35°

5.0

Radiographic Imaging Table with the following features:

5.1

Four-way floating ,non-transparent flat tabletop with electro-magnetic brakes for tabletop movements and motorized height adjustment.

5.2

Foot switch(es) built-in to the table base for brake release of tabletop movements and height adjustment,

5.3

Safety interlock/device to stop tabletop movement when footswitch is inadvertently stepped on 

5.4

Collision prevention device to stop tabletop vertical  movement if lowers onto any obstruction

5.5

Table-top movement range

5.6

Longitudinal: ±325mm, Transverse: no less than ±110 mm, Size of the table is 2000x760mm. Patient weight carrying capacity of tabletop is 200kg  Lowest table height,

5.7

Integrated with the CCD detector 

5.8

Equipped with a removable, oscillating grid ,grid ratio 10:1 

5.9

Bucky assembly have a vertical movement range of 470mm~1468mm

6.0

The table detector shall have the following features:

6.1

Data acquisition is 14 bits per pixel

6.2

Image resolution is at least 4.6 lp/mm,7 seconds imaging

Product Name:
High Frequency Mobile C arm x ray  
Item No: TDM8160A